Abstract
A simple impedimetric label-free immunosensor was developed for the specific and sensitive detection of mucin 4 (MUC 4) protein by using graphite based screen printed electrodes modified with an aryl diazonium salt or compound (p- aminophenylacetic acid) for the immobilization of antibody anti-MUC4 via amidic bond. Electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV) were used in order to characterize and optimize the electrografting process. The parameters involved in each step of the immunosensor design were optimized. The performance of the immunoassay in terms of sensitivity, reproducibility and selectivity was studied.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.