Abstract

There are three principal magnetic fields in magnetic resonance imaging (MRI) that may interact with medical implants. The static field will induce force and torque on ferromagnetic objects. The pulsed gradients are of audio frequency and the implant may concentrate the induced currents, with a potential for nerve stimulation or electrical inference. The currents induced in the body by the radio frequency (RF) field may also be concentrated by an implant, resulting in potentially dangerous heating of surrounding tissues. This paper presents basic information about MRI interactions with implants with an emphasis on RF-induced heating of leads used for deep brain stimulation (DBS). The temperature rise at the electrodes was measured in vitro as a function of the overall length of a DBS lead at an RF frequency of 64 MHz. The maximal temperature rise occurred for an overall length of 41 cm. The method of moments was used to calculate the current induced in the lead. From the induced currents, the RF power deposition near the electrodes was calculated and the heat equation was used to model the temperature rise. The calculated temperature rises as a function of lead length were in good agreement with the measured values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.