Abstract

MR microscopy on intact rats using a single surface coil is reported. The use of the same coil for excitation and detection is experimentally very simple and results in a smaller field-of-view than the conventional setup using homogeneous excitation (whole-body coil or a large surface coil) and a small receiver surface coil. Images at 4.7 T with 60 X 60-microns pixels and an 800-microns slice thickness have been obtained from the eye and ankle joint of a living rat with good anatomical definition. This corresponds to a four- to eightfold reduction in voxel size as compared to high-resolution images obtained with a 30-mm resonator probe. Many of the basic MRI experiments (spin-echo, FLASH, chemical-shift selective MRI) may be carried out in a straightforward way with this one-coil setup. Sequences relying on a homogeneous flip angle distribution require a whole-body excitation coil.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.