Abstract

Magnetic resonance (MR) images of the pelvis obtained with an endorectal coil have improved resolution relative to that of body coil images; however, they have limited spatial coverage. The sensitivity profile can be improved by using the endorectal coil combined with an external anterior coil as part of a multicoil array. If each coil is connected to a separate receiver, the individual images can be combined to provide an optimal signal-to-noise ratio (S/N). The authors have investigated a 5-inch (12.7-cm) surface coil or a two-coil array as the anterior coil. In both of these configurations, the S/N is improved relative to that of the endorectal coil alone at distances greater than 3 cm from the endorectal coil. The anterior two-coil array provides more lateral coverage. The improved spatial coverage available with this technique expands potential clinical applications of endorectal MR imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.