Abstract

Recently developed techniques such as the multiple acquisition with variable resonance image combination and slice encoding for metal artifact correction techniques have improved the ability of clinical magnetic resonance scanners to image near metal implants. These sequences are based on fast spin echo sequences which preclude detection of short T(2) tissues such as tendons, ligaments, and cortical bone. Ultrashort echo time sequences have the potential to detect signals from these tissues. In this study, we investigate the potential of combining ultrashort echo time with multiple acquisition with variable resonance image combination to image short T(2) musculoskeletal tissues adjacent to metallic implants. Different radio frequency excitation pulse types and spectral binning strategies were studied. We found that ultrashort echo time-multiple acquisition with variable resonance image combination sequences were able to significantly reduce typical artifacts near metal, as well as detect very short T(2) signals that are usually not visualized using clinical pulse sequences.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.