Abstract
To compare the results and reproducibility of two MR-based methods of measuring the cerebrovascular response (CVR). In eight volunteers, CVR was assessed with two MR-based methods upon a challenge with acetazolamide. CVR was assessed by measuring changes in total cerebral blood flow (TCBF) using phase contrast (PC) MRI, and by measuring perfusion MRI. To assess reproducibility the measurements were repeated after 1 week. The average CVR with the PC-MRI method was 46% (SD = 16%), and for perfusion MR the measured CVR was 44% (SD = 16%). The coefficient of variation (COV) for PC-MRI was 28%, while perfusion MR had a COV of 26%. The limits of agreement between the two methods were -49% and 45%, demonstrating a lack of agreement between the two methods in terms of CVR estimation. CVR estimates based on PC-MRI and perfusion MRI showed reproducibility but a lack of agreement in healthy volunteers. This lack of agreement can be attributed to the different aspects of the CVR reflected by these methods: TCBF reflects changes in CBF, whereas our perfusion MRI method reflects cerebral blood volume (CBV).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.