Abstract

Two hundred and seventy mice were evaluated at three exposure durations (5, 10 and 20 min) and at six peak acoustic pressure levels (0, 65, 80, 87, 100 and 145 kPa) with 15 mice per exposure condition, at an ultrasonic frequency of 30 kHz modulated at 120 Hz. Threshold acoustic pressure levels for hemorrhage in mouse lung exposed for the three exposure durations appear to be in the range of 100 kPa. There did not appear to be a strong dependency on exposure duration. When compared to a study with mice in the megahertz frequency range (Child et al. 1990), it appeared that the threshold values followed a square root of frequency dependency, suggesting that the concept of the Mechanical Index (AIUM/NEMA, 1992), although developed for pulsed ultrasound conditions, may be extended to frequencies well below the diagnostic ultrasound frequency range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.