Abstract

The aim of this work is to analyze the variability in manual measurements of motor unit action potential (MUAP) duration and to evaluate the effectiveness of well-known algorithms for automatic measurement. Two electromyographists carried out three independent duration measurements of a set of 240 MUAPs. The intraexaminer and interexaminer variabilities were analyzed by means of the Gage Reproducibility and Repeatability method. The mean of the three closest manually marked positions was considered the gold standard of the duration markers positions (GSP). The results of four well-known automatic methods for estimating MUAP duration were compared to the GSP. Manual measurements of duration showed a lot of variability, with the combined intraoperator and interoperator variability greater than 30%. The greatest difference between manual positions was 11.2 ms. The mean differences between the GSP and those obtained with the four automatic methods ranged between 0.6 and 8.5 ms. Both manual and automatic measurements of MUAP duration show a high degree of variability. More precise methods are needed to improve the accuracy and reliability of the estimates of this parameter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.