Abstract
In magnetic resonance (MR) cine imaging applications, it is common that only a fraction of the field of view is used for delineating dynamic features whereas surrounding static structures are merely served as landmarks for three-dimensional orientation. This paper presents a method of using motion-selective encoding for fast cine imaging. By the use of a simple RF saturation pulse, the imaging field of view was divided into two areas, one with motion and the other without. This allowed the static material to be imaged by fast ungated acquisition and dynamic structures by reduced phase encoding. Issues related to phase error correction were addressed and the proposed algorithm was validated using images acquired from five normal subjects. The saving in imaging time was dependent on the size of the moving structures and a 35% reduction in imaging time did not cause visually noticeable changes in image quality. The application of the technique is limited to cine images with large surrounding static structures and a theoretical analysis of maximum savings in imaging time in relation to the size of the moving structures within the image field of view is provided. Magn Reson Med 42:430-435, 1999.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.