Abstract

PurposeSubject motion is a major source of image degradation for functional MRI (fMRI), especially when using multishot sequences like three‐dimensional (3D EPI). We present a hybrid radial‐Cartesian 3D EPI trajectory enabling motion correction in k‐space for functional MRI.MethodsThe EPI “blades” of the 3D hybrid radial‐Cartesian EPI sequence, called TURBINE, are rotated about the phase‐encoding axis to fill out a cylinder in 3D k‐space. Angular blades are acquired over time using a golden‐angle rotation increment, allowing reconstruction at flexible temporal resolution. The self‐navigating properties of the sequence are used to determine motion parameters from a high temporal‐resolution navigator time series. The motion is corrected in k‐space as part of the image reconstruction, and evaluated for experiments with both cued and natural motion.ResultsWe demonstrate that the motion correction works robustly and that we can achieve substantial artifact reduction as well as improvement in temporal signal‐to‐noise ratio and fMRI activation in the presence of both severe and subtle motion.ConclusionWe show the potential for hybrid radial‐Cartesian 3D EPI to substantially reduce artifacts for application in fMRI, especially for subject groups with significant head motion. The motion correction approach does not prolong the scan, and no extra hardware is required. Magn Reson Med 78:527–540, 2017. © 2016 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

Highlights

  • Three-dimensional (3D) echo planar imaging (EPI) is attractive for high-resolution functional MRI, primarily due to gains in signal-to-noise ratio (SNR) efficiency as a result of using volumetric excitation and 3D Fourier encoding [1,2,3,4]

  • EPI planes are rotated in a golden-ratio angle scheme [13] about the phase-encoding axis. This sequence has a number of benefits for fMRI: Key words: functional MRI; 3D EPI; radial-Cartesian EPI; motion correction; self-navigated; golden angle

  • We focus on investigating the potential of Trajectory Using Radially Batched Internal Navigator Echoes (TURBINE) fMRI for motion correction

Read more

Summary

Introduction

Three-dimensional (3D) echo planar imaging (EPI) is attractive for high-resolution functional MRI (fMRI), primarily due to gains in signal-to-noise ratio (SNR) efficiency as a result of using volumetric excitation and 3D Fourier encoding [1,2,3,4]. The most commonly used 3D trajectory for fMRI is Cartesian multishot 3D EPI [1], for which readout planes are parallel. EPI planes are rotated in a golden-ratio angle scheme [13] about the phase-encoding axis. This sequence has a number of benefits for fMRI: Key words: functional MRI; 3D EPI; radial-Cartesian EPI; motion correction; self-navigated; golden angle

Objectives
Methods
Results
Conclusion