Abstract

Insulin assays are utilized in various clinical scenarios, including the assessment of insulin therapy compliance or of suspected insulin overdose. In an interpretative exercise carried out by UK National External Quality Assessment Service (NEQAS), serum sent to the participating laboratories was spiked with 30 pmol/L of the short-acting insulin analogue Human Actrapid. Only two out of 24 participant laboratories had sufficient assay cross-reactivity with Actrapid to interpret the results as suggestive of insulin administration. The development of specific insulin assays has led to deterioration in the ability to detect non-compliance or overdose with recombinant insulin treatment. Clinicians should be aware of this significant limitation, which could lead to misdiagnosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.