Abstract

We investigated left ventricular ejection fraction and new left ventricular wall motion abnormality before pacemaker implantation and at follow-up and mortality at long-term follow-up in 81 patients with implantable cardioverter defibrillators treated with biventricular pacing and in 80 patients with implantable cardioverter defibrillators treated with DDDR-70 right ventricular pacing. At 23-month follow-up, the left ventricular ejection fraction decreased from 36% to 30% in patients treated with right ventricular pacing and increased at 38-month follow-up from 35% to 40% in patients treated with biventricular pacing (P < 0.001). New left ventricular wall motion abnormality developed at 23-month follow-up in 23 of 80 patients (29%) treated with right ventricular pacing and at 38-month follow-up in 7 of 81 patients (9%) treated with biventricular pacing (P < 0.005). Twenty-two of 80 patients (28%) treated with right ventricular pacing died at 45-month follow-up and 8 of 81 patients (10%) treated with biventricular pacing died at 53-month follow-up (P < 0.01).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call