Abstract

We studied the effect of varying impaction force, repeated impactions, and fluid contamination on the disassembly strength of Morse-type tapers in 4 commercially available, modular femoral total hip components. The effect of varying techniques of taper assembly on the distraction force was studied. Our results show a reproducible and linear relationship between the taper impaction force and the disassembly force. The force necessary to separate the taper for a given impaction force varied, however, among manufacturers. Repeated impactions added little strength, and we found that when multiple impactions of varying force are used, the strength is roughly equivalent to the expected strength from the single strongest blow. Fluid contamination at the taper interface had unpredictable effects on taper strength. Copyright 2002, Elsevier Science (USA). All rights reserved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.