Abstract

Transmission crystal spectrometers (TCS) are used on many laser facilities to record hard X-ray spectra. During experiments, signal recorded on imaging plates is often degraded by a background noise. Monte-Carlo simulations made with the code GEANT4 show that this background noise is mainly generated by diffusion of MeV electrons and very hard X-rays. An experiment, carried out at LULI2000, confirmed that the use of magnets in front of the diagnostic, that bent the electron trajectories, reduces significantly this background. The new spectrometer SPECTIX (Spectromètre PETAL à Cristal en TransmIssion X), built for the LMJ/PETAL facility, will include this optimized shielding.