Abstract

The purpose of this work is to assess 8 detectors variation performance for output factor (OF) determination, in a clinical 6 MV Varian Clinac 2100C stereotactic radiosurgery mode for cone irradiation using Monte Carlo simulation as reference. We study 10 cones with diameters between 30 and 4 mm. The evaluated detectors were ionization chambers: pinpoint and pinpoint 3D (PTW), diodes: SRS, P and E (PTW), Edge (Sun Nuclear), microdiamond (PTW) and EBT3 radiochromic films (Ashland). The OF were normalized with the 30 mm cone and compared to Monte Carlo (MC). For the 8 detectors, the OF measurements reproducibility was very high: r = 0.99 and p 7% for smaller cones. Microdiamond slightly overestimates OF, 3.7% for all the cones and EBT3 film is the closest to MC with maximum difference ±1% whatever the cone size. Film is the more accurate detector for OF determination of stereotactic cones but it is restrictive to use. Pinpoints and diodes, respectively due to inappropriate size of sensitive volume and composition do not seem appropriate without corrective factors below 10 mm diameter cone. Microdiamond appears the best detector for all cones despite its sensitive volume size.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.