Abstract

The purpose of this study was to investigate the characteristics of electron contamination and Output Factors (OFs) from Varian Trilogy Clinac iX 6 MV photon beam at small field sizes. EGSnrc Monte Carlo (MC) code system was used to model the photon beam for this Linear Accelerator (Linac) head and analyze the electron contamination and OFs from this treatment head. The electron contamination was analyzed for field sizes of 1 × 1, 2 × 2, 3 × 3, 4 × 4, and 5 × 5 cm2. The number of electron contamination increases with increasing field sizes, but the maximum energy of the electron contamination stays constant (at around 1.87 MeV for each field size). The contaminants contribute to the dose at the surface of the water phantom (1–5 cm from the surface) for field size 4 × 4 and 5 × 5 cm2 and this dose decreases with depth. The OFs are simulated by EGSnrc code system and have a good agreement with measurement (deviation 3.45, 1.76, and 0.86 for field of 2, 3 and 4, respectively). This study presented that MC methods have great potential to accurately predict the electron contamination and OFs for 6 MV photon beam.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.