MADPET-4 is a high resolution PET insert under development for use in a 7-T MR. To fully exploit the capabilities of the insert, a good understanding of the physical interactions, which take place in the active and passive components of the insert, is necessary. The goal of this study was to investigate the effects of different physical interactions in an accurate model of MADPET-4 using Monte Carlo (MC) simulations. The main focus of the study was on the impact of the different active and passive components of the system on the amount of random and scattered events, including scattering in the passive components and between the crystals. The influence of the low energy threshold (50-350 keV) and different geometrical conditions in the coincidence sorting process was of particular interest. The effect of including triple coincidences was also considered in the present study. Results showed that the maximum sensitivity achieved at the minimum energy threshold (50 keV), using a geometrical condition of 4-9 sectors, was ~3.43% without passive components and dropping to ~1.36% including the passive components. Interestingly, the minimum noise ratio was obtained with a geometrical condition of 9 sectors difference at an energy threshold of 50 keV. Including triple coincidences showed no benefit for energy thresholds above 250 keV. However, low energy thresholds provided a relative gain in sensitivity, reaching 15-20% for 200 keV and a maximum of 30-40% at 50 keV.

Highlights

  • MADPET-4 is a high resolution PET insert under development for use in a 7-T MR

  • The goal of this study was to investigate the effects of different physical interactions in an accurate model of MADPET-4 using Monte Carlo (MC) simulations

  • The main focus of the study was on the impact of the different active and passive components of the system on the amount of random and scattered events, including scattering in the passive components and between the crystals

Read more

Summary

Introduction

MADPET-4 is a high resolution PET insert under development for use in a 7-T MR. To fully exploit the capabilities of the insert, a good understanding of the physical interactions, which take place in the active and passive components of the insert, is necessary. Monte Carlo study of scattered and random coincidences for MADPET-4 Negar Omidvari1*, Jorge Cabello1, Florian Schneider1, Stephan Paul2, Sibyle I Ziegler1

Objectives
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.