Abstract

Monte Carlo methods have been used to simulate the scattering of x rays in polystyrene and water phantoms. In particular, the ratio of the scattered to total x-ray fluence (scatter fraction) has been calculated for monoenergetic x-ray beams in the energy region relevant to diagnostic radiology and nuclear medicine (30-660 keV). Simulations have been made for representative values of the pertinent geometrical factors; phantom thickness from 5 to 21 cm, x-ray beam diameters of 10 and 25 cm, and scatterer-to-image-plane separations from 0 to 20 cm. As a function of x-ray energy, the scatter fraction was found to vary slowly between 30 and 100 keV, and to decrease between 100 and 660 keV. The present results were generated with a special transport code which included the effects of special geometries and the response of the x-ray detector. With the inclusion of these effects, the results resolved inconsistencies and showed good agreement with previous measured and calculated data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.