Abstract

The new manganese bath facility of the Laboratoire National Henri Becquerel has been modeled by using three Monte-Carlo codes: MCNPX, GEANT4, and FLUKA, in order to determine the correction factors needed in the neutron source calibration process. The most realistic source geometry has been determined, and the most reliable cross sections library has been chosen. The models were compared, and discrepancies between the codes have been pointed out. Potential causes of deviations between results were assessed and discussed using additional models. Finally, an experimental process is proposed to validate the accuracy of the different codes and their abilities in simulating the neutron capture by the manganese bath.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.