Abstract

To use MC simulations and phantom measurements to investigate the dosimetry of a kilovoltage x-ray beam from an IR fluoroscope to deliver low-dose (0.3-1.0Gy) radiotherapy to the lungs. PENELOPE was used to model a 125kV, 5.94mm Al HVL x-ray beam produced by a fluoroscope. The model was validated through depth-dose, in-plane/cross-plane profiles and absorbed dose at 2.5-, 5.1-, 10.2- and 15.2-cm depths against the measured beam in an acrylic phantom. CT images of an anthropomorphic phantom thorax/lungs were used to simulate 0.5Gy dose distributions for PA, AP/PA, 3-field and 4-field treatments. DVHs were generated to assess the dose to the lungs and nearby organs. Gafchromic film was used to measure doses in the phantom exposed to PA and 4-field treatments, and compared to the MC simulations. Depth-dose and profile results were within 3.2% and 7.8% of the MC data uncertainty, respectively, while dose gamma analysis ranged from 0.7 to 1.0. Mean dose to the lungs were 1.1-, 0.8-, 0.9-, and 0.8- Gy for the PA, AP/PA, 3-field, and 4-field after isodose normalization to cover∼95% of each lung volume. Skin dose toxicity was highest for the PA and lowest for the 4-field, and both arrangements successfully delivered the treatment on the phantom. However, the dose distribution for the PA was highly non-uniform and produced skin doses up to 4Gy. The dose distribution for the 4-field produced a uniform 0.6Gy dose throughout the lungs, with a maximum dose of 0.73Gy. The average percent difference between experimental and Monte Carlo values were -0.1% (range -3% to +4%) for the PA treatment and 0.3% (range -10.3% to +15.2%) for the 4-field treatment. A 125kV x-ray beam from an IR fluoroscope delivered through two or more fields can deliver an effective low-dose radiotherapy treatment to the lungs. The 4-field arrangement not only provides an effective treatment, but also significant dose sparing to healthy organs, including skin, compared to the PA treatment. Use of fluoroscopy appears to be a viable alternative to megavoltage radiation therapy equipment for delivering low-dose radiotherapy to the lungs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call