Abstract

The purpose of this article is to discuss the method of determining the wall correction factor of graphite cavity ionisation chamber. The method is the ‘equal effect wall thickness’ simulation and direct calculation of the wall correction factor of the std10cm3 and std30cm3 spherical graphite cavity ionisation chamber in the national air-specific kinetic energy reference group (National Institute of Standards and Technology NIST), respectively, in the NIST10cm3 and NIST30cm3 spherical graphite cavity ionisation chamber. The calculated results are in agreement with the values released by NIST in 0.1%. Conclusions of the equivalent wall thickness simulation method and the direct calculation simulation method meet each other's requirements, which provide a new idea for the determination of wall correction factor.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.