Abstract
Topographic contrast of secondary-electron (SE) images in a scanning ion microscope (SIM) using a focused gallium (Ga) ion beam is investigated by Monte Carlo simulation. The SE yield of heavy materials, in particular, due to the impact of 30 keV Ga ions increases much faster than for the impact of electrons at < or =10 keV as a function of the angle of incidence of the primary beam. This indicates the topographic contrast for heavy materials is clearer in a SIM image than in a scanning electron microscope (SEM) image; for light materials both contrasts are similar to each other. Semicircular rods with different radii and steps with large heights and a small wall angle, made of Si and Au, are modeled for comparison with SE images in SEM. Line profiles of the SE intensity and pseudo-images constructed from the profiles reveal some differences of the topographic contrast between SIM and SEM. We discuss not only the incident-angle effect on the contrast, but also the effects of re-entrances of primary particles and SEs to the neighboring surface, the effect of a sharp edge on the sample surface, and the effects of pattern size and beam size.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.