A Monte Carlo study of the transient response of single photon absorption in X-ray pixel detectors is presented. The simulation results have been combined with Monte Carlo simulation of the X-ray photon transport and absorption, and used to estimate the image properties of a detector system, including the pixel array and readout electronics. The study includes several different simulation challenges, such as full band Monte Carlo simulation of charge transport in large devices ( 300 μm×100 μm ), modelling of three-dimensional electrostatic effects using cylindrical coordinates, Monte Carlo simulation of photon transport and absorption, and a system level Monte Carlo simulation of the entire pixel detector and readout.