Abstract
We report results of scatter simulations for both point and distributed sources of 99mTc in symmetrical non-uniform attenuating media. The simulations utilized Monte Carlo techniques and were tested against experimental phantoms. Both point and ring sources were used inside a 10.5 cm radius acrylic phantom. Attenuating media consisted of combinations of water, ground beef (to simulate muscle mass), air and bone meal (to simulate bone mass). We estimated/measured energy spectra, detector efficiencies and peak height ratios for all cases. In all cases, the simulated spectra agree with the experimentally measured spectra within 2 SD. Detector efficiencies and peak height ratios also are in agreement. The Monte Carlo code is able to properly model the non-uniform attenuating media used in this project. With verification of the simulations, it is possible to perform initial evaluation studies of scatter correction algorithms by evaluating the mechanisms of action of the correction algorithm on the simulated spectra where the magnitude and sources of scatter are known.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Radiation Applications & Instrumentation. Part A, Applied Radiation & Isotopes
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.