Abstract

Transmission crystal spectrometers (TCS) are used on many laser facilities to record hard X-ray spectra. During experiments, signal recorded on imaging plates is often degraded by a background noise. Monte-Carlo simulations made with the code GEANT4 show that this background noise is mainly generated by diffusion of MeV electrons and very hard X-rays. An experiment, carried out at LULI2000, confirmed that the use of magnets in front of the diagnostic, that bent the electron trajectories, reduces significantly this background. The new spectrometer SPECTIX (Spectromètre PETAL à Cristal en TransmIssion X), built for the LMJ/PETAL facility, will include this optimized shielding.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.