Abstract
A new method for prediction of gas chromatographic retention times and peak half widths is based on the renewal theory. The only requirements are the heats of vaporization of the compounds to be separated and one calibration measurement. With this data, retention times and peak half widths can be predicted for isothermal as well as temperature-programmed gas chromatography. For the separation of non-polar substances on non-polar stationary phases the prediction error for retention times is approx. 1–2%. First simulations of polar molecules and polar stationary phases indicate that this method is also applicable in these cases but some extension will be required.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.