Abstract

This study aimed to obtain the absorbed dose distribution in a heterogeneous head simulator object where a brain tumor of 2 cm in diameter was located. Materials equivalent to the head and tumor were created. For head and tumor simulation, elemental composition (H, C, N, O, and others) and mass density ρ respectively were used. X-ray spectra of 6, 10 and 18 MV were used for the exposures, projecting a field of 3×3 cm2 to the tumor, with an isocentric source distance of 100 cm. Monte Carlo simulations were done with PENELOPE v.2008 code. The results show that the maximum dose to the skin is 32 %, 20 % and 14 %, the maximum dose to the skull is 88 %, 74 % and 62 %, and the maximum dose to the tumor is 62 %, 67% and 73 %, for energies of 6, 10 and 18 MV respectively. The maximum dose received by the skin and skull tissues decreases with increasing energy, while the dose in the tumor increases with increasing energy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.