Abstract

The purpose of this study is to investigate the feasibility of a novel indirect flat panel detector (FPD) system for removing scatter radiation. The substrate layer of our FPD system has a Pb net-like structure that matches the ineffective area and blocks the scatter radiation such that only primary X-rays reach the effective area on a thin-film transistor. To evaluate the performance of the proposed system, we used Monte Carlo simulations to derive the scatter fraction and contrast. The scatter fraction of the proposed system is lower than that of a parallel grid system, and the contrast is superior to that of a system without a grid. If the structure of the proposed FPD system is optimized with respect to the specifications of a specific detector, the purpose of the examination, and the energy range used, the FPD can be useful in diagnostic radiology.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.