Abstract

In this study, the geometry of the head of VARIAN CLINAC 2100C linear accelerator lodged at Blida Anti-Cancer Centre in Algeria has been modeled using GATE/GEANT4 version 7.2. The model developed contains the majority of the components of the treatment head operating in 18 MV, such as: the target, the primary collimators, the flattening filter, the secondary collimators and the jaws. Using this model, associated with a water phantom of type RFA 200 with dimensions 63.8 cm × 40 cm × 44 cm positioned at a Source-to-Surface-Distance (SSD) of 100 cm, the interactions of the 18 MV radiations in water were simulated. The photon energy spectrum, the Percentage Depth Dose (PDD) and the beam quality specifiers such as D10 (%), dmax(cm), d80(cm), D20/D10 were calculated for field size of 10 cm x 10 cm utilizing the phase space technique. The results obtained in this study are in agreement with those obtained in our previous work using MCNP5 code.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.