Abstract

An MCNP model of a pair of planar HPGe detectors (designated as: detector 3 and detector 4), that are used for routine lung counting at AECL, was developed. The model was benchmarked against experimental results, where a multi-line 152Eu source was counted in several different geometrical arrangements. The best agreement for both detectors was achieved when side and back dead layers (of both detectors) were quadrupled, with respect to the ones quoted by their manufacturer (Canberra). In the case of detector 4, the agreement between simulated and measured spectra was within 4%, throughout the whole γ-spectrum, spanning 70–1600keV. The same was true for detector 3 at the lower end of the γ-spectrum. However, at the high end of the γ-spectrum, the agreement was within 7% and 12% for 152Eu γ-lines at 778.9 and 1408.01keV.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.