Abstract

A Monte Carlo method has been used to model the measument of cisplatin uptake with in vivo X-ray fluorescence. A user-code has been written for the EGS4 Monte Carlo system that incorporates linear polarisation and multiple element fluorescence extensoions. The yield of fluorescent photons to the mainly Compton scattered background is computed for our detector arrangement. The detector consists of a mutually orthogonal arrangement of X-ray tube, aluminium polariser and high purity germanium scintillation detector. The influence of tube voltage on the minimum detectable concentration is modelled for 100 through 150 kVp X-radiation. The code is able to predict absorbed dose to the patient which will influence the optimal choice of tube voltage. The influence of alterations to collimator design and scatterer onstruction can also be examined. A minimum detectable concentration of 50 ppm is determined from measurements with a 115 kVp X-ray source and a 615 ppm cisplatin sample in a water phantom.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.