The Monte Carlo method enables accurate dose calculation for radiotherapy treatment planning and has been implemented in some commercial treatment planning systems. Unlike conventional dose calculation algorithms that provide patient dose information in terms of dose to water the Monte Carlo method calculates the energy deposition in different media and expresses dose to medium. This paper discusses the differences in dose calculated using water with different electron densities and that calculated for different biological media and the clinical issues in reporting dose using dose to water and dose to medium. A recommendation is made based on our studies to report dose in medium for radiotherapy treatment plan evaluation and for radiotherapy treatment outcome analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.