Abstract

This study aims at the Monte Carlo (MC) commissioning of Varian Clinac 2100C medical linear accelerator (LINAC) for 6 MV photon beam. The MC-based EGSnrc user sub codes BEAMnrc and DOSXYZnrc were used for accelerator head modeling and dose calculations in water phantom respectively. As the first step in the commissioning of LINAC, optimum incident electron beam energy was investigated. For that purpose, lateral beam profiles (at different depths) and percentage depth dose curves were acquired for different field sizes (5 ×5, 10 × 10, 20 × 20 and 30 × 30 cm2). By the comparison of simulated and measured dose profiles, the optimum incident electron beam energy was found to be 5.8 MeV using x2 test. Lateral beam profiles and depth dose curves agreed well with reference data. However, lateral beam profiles are found to be more sensitive than depth doses to changes in incident electron beam energy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.