Abstract

The application of cavity-ionization chambers to the standadization of 60Co gamma-ray beams, in terms of exposure, requires that the specific ionization of air Jg, be corrected for the attenuation and scatter of the incident rays by the wall, central electrode, and supporting stem of the chamber. A Monte Carlo photon-electron transport code has been developed for the purpose of calculating this correction for spherical and cylindrical chambers. The code has been applied to a spherical graphite chamber having dimensions typical of the chambers used by the NBS, the calculated wall-correction factor is in close agreement with the average of the NBS factors which were determined experimentally. The code was also used to calculate Aeq, which is central to the determination of tissue-air ratios. The calculated value, 0.989 +/- 0.003, is very close to the generally accepted value, 0.985.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.