Abstract

Spectra of calibration sources and X-ray beams were measured with a cadmium telluride (CdTe) detector. The response function of the detector was simulated using the GEANT4 Monte Carlo toolkit. Trapping of charge carriers were taken into account using the Hecht equation in the active zone of the CdTe crystal associated with a continuous function to produce drop of charge collection efficiency near the metallic contacts and borders. The rise time discrimination is approximated by a cut in the depth of the interaction relative to cathode and corrections that depend on the pulse amplitude. The least-squares method with truncation was employed to unfold X-ray spectra typically used in medical diagnostics and the results were compared with reference data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.