Abstract

For accurate evaluation of air-kerma strength, Sk, of 169Yb and 60Co brachytherapy sources, the present study reports Monte Carlo (MC) based corrections for (1) room scatter, and (2) departure from constant room scatter for rooms of various sizes. Correction for exponential attenuation of effective primary in air is also reported for the above sources.Values of Sk per contained mCi, Sk/Ac, predicted by MC calculations for 169Yb source (model X1267) with and without Ti K x-rays are 1.302 ± 0.03% (this value is in excellent agreement with the published value reported by Piermattei et al) and 1.260 ± 0.03% cGy cm2 h−1 mCi−1 respectively, and in the case of Cathetron 60Co source the value of Sk/Ac is 11.015 ± 0.01% cGy cm2 h−1 mCi−1.It is observed that depending upon the position of the source with respect to the surrounding concrete scattering surfaces and set of d values, the assumption of constant room scatter has resulted in overestimation of Sk that varied between 0.30% and 1.5% for the 169Yb source and only between 0.10% and 0.20% for the 60Co source.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.