Abstract

The principal methods of employing double-crystal reflection for harmonic rejection are briefly reviewed. The effect of using non-equal d spacings (dispersion) is considered. A new type of perfect-crystal monochromator consisting of a monolithic combination of a pair of grooved crystals a and b is described. The monochromaticity (degree of harmonic suppression) M(β) is tuned by tilting groove a with respect to groove b by an appropriate angle β. The tilting is achieved in a simple way through a pair of helical springs. With three reflections in the first and two reflections in the second groove M(β0) values between about 300 and 100 have been measured for X-rays at energies ranging from 6.72 to 9.86 keV and at a bandwidth ΔE/E≃1.5%. β0 is the tilting angle for which M(β) is optimized. Once aligned the monochromator is a very simple unit, which can be mounted on any standard goniometer head and which can be used for both X-rays and thermal neutrons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.