Abstract

Monolithic capillary columns based on divinylbenzene were synthesized using different alcanols as porogens. Prepared columns were tested in separation of polystyrene standards according to their molar mass (MM) and were characterized by corresponding calibration graphs. It was demonstrated that a decrease of alcanol chain length from dodecanol to octanol resulted in a decrease of column permeability and in an improved column ability to separate polystyrene standards. In contrast, removing a good solvent from porogen mixture results in an increase of column permeability and in a decrease of column performance toward polystyrene standards. Optimized synthetic conditions included porogen composed of nonanol and toluene or mesytilene, and the column prepared with this porogen was capable of separating a mixture of 14 polystyrene standards with MM ranged from several millions to oligomers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.