Abstract
Poly(styrene-co-divinylbenzene)-based monolithic capillaries of an inner diameter of 200 mum and a length of 2-5 mm have been used to construct Ca2+-, Ag+-, and Na+-selective electrodes. The membranes consist of a solution of ionophore and ion exchanger in bis(2-ethylhexyl) sebacate or 2-nitrophenyl octyl ether, which are used as plasticizers in conventional PVC-based membranes. With capillaries of low porosity, the potentiometric responses down to 10(-8)-10(-9) M solutions do not depend on the composition of the internal solution, which indicates a strong suppression of transmembrane ion fluxes. Thus, no tedious optimization of the inner solution is required with monolith ISEs. The lower detection limits of Ag+- and Ca2+-ISEs are comparable to the best ones obtained earlier with optimized inner solutions. Additionally, a monolithic Na+-selective ISE has been obtained exhibiting a lower detection limit of 3 x 10(-8) M Na+. With monolithic capillaries of higher porosity and fused-silica GC capillaries, the transmembrane flux effects are noticeable but still significantly smaller than with conventional PVC membranes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.