Abstract

A source has been developed which produces a pure spectrum of Lyman-α radiation (1215.7 Å). This source incorporates a wavelength selective filter and an rf-excited helium-filled lamp containing a mixture of uranium and uranium hydride in a sidearm. The uranium serves as a getter to eliminate atmospheric contaminants, and the uranium hydride, when heated, supplies H2 in a reproducible manner. The filter consists of a flowing-oxygen cell and a narrowband interference filter. The distinctive advantage of this device is that radiation in the VUV at a well-defined wavelength is obtained without the use of a monochromator. Characteristics of the source and measurements of the irradiance of the spectral line are given for a typical lamp. The irradiance and spectral purity are seen to be not strongly dependent on oxygen flow.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.