Abstract

An improved monochromatic radiation source with spectral width 4 nm based on a supercontinuum laser and double monochromator is introduced into a unit based on an absolute cryogenic radiometer in order to increase the precision of measurement of spectral sensitivity in the range 0.9–1.6 μm. A feedback system that supports stabilization of the power of monochromatic radiation with standard deviation 0.025% is developed. The power of optical radiation in the plane of the detector and/or power of the absolute cryogenic radiometer varies from 0.1 to 1.5 mW. The spectral distribution of the power of the newly developed source in different operating regimes of the supercontinuum laser is presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.