Abstract
This paper shows the design of a fiber-based sensor for living activities in human body and the results of a laboratory evaluation carried out on it. The authors have developed a device that allows for monitoring the vibrations of human body evoked by living activities--breathing and cardiac rhythm. The device consists of a Bragg grating inscribed into a single mode optical fiber and operating on a wavelength of around 1550 nm. The fiber Bragg grating (FBG) is mounted inside a pneumatic cushion to be placed between the backrest of the seat and the back of the monitored person. Deformations of the cushion, involving deformations of the FBG, are proportional to the vibrations of the body leaning on the cushion. Laboratory studies have shown that the sensor allows for obtaining dynamic strains on the sensing FBG in the range of 50-124 μ strain caused by breathing and approximately 8.3 μstrain induced by heartbeat, which are fully measurable by today's FBG interrogation systems. The maximum relative measurement error of the presented sensor is 12%. The sensor's simple design enables it to be easily implemented in pilot's and driver's seats for monitoring the physiological condition of pilots and drivers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.