Abstract

Pharmaceutical manufacturing utilizes solvents at different stages of production. Some of the harmful solvent residuals may be retained in the final product; therefore, they need to be monitored for quality control and to meet the regulation requirement. Here, a novel method capable of rapidly analyzing residual solvents in pharmaceutical products was developed using a compact-portable gas chromatography with a photoionization detector (GC-PID). The method consists of modified Tedlar® bag sampling, online pre-concentration, separation of volatiles by miniaturized GC, and micro-PID detection. The method detection limits of selected residual solvents were in the range of 26.00 – 52.03 pg/mL which is much lower than the pharmaceutical compliance concentration limits. Limits of detection > 520 pg of analyte per grams of sample was also determined for the over-the-counter drugs. The method performance showed rapid speed (5 min), linear calibration (r2 < 0.99), and repeatable retention time (RSD < 0.4 %). Direct analysis of residual solvents in solid samples was conducted without the need for complex sample preparation. The method validation using over-the-counter pharmaceutical products yielded excellent accuracy (recovery > 91.2 %) and precision (RSD < 6.5 %) for the selected residual solvents, including 1,4-dioxane, benzene, chlorobenzene, cyclohexane, xylenes, and toluene. This portable and rapid method could be deployed during the pharmaceutical drug manufacturing processes for quality control.