Abstract

This study was aimed at validating the use of a custom-made wireless pulse oximeter in freely moving lambs, using radiotelemetry transmission. First, measurements obtained simultaneously using the new, wireless oximeter and a standard commercially-available pulse oximeter (Nonin 8500) were compared in five lambs during 5min episodes of normoxia, hypoxia and hyperoxia. Correlation between the two oximeters for both SpO(2) and heart rate was very good, regardless of oxygenation conditions. Secondly, the capabilities of our device were assessed during more than 45h of polysomnographic recordings in seven lambs. According to the plethysmographic pulse waveform, reliable SpO(2) values were obtained in more than 85% of recording time. Multiple decreases in SpO(2) were readily observed after spontaneous apneas in preterm lambs. It is concluded that our wireless pulse oximeter performs as reliably as a standard pulse oximeter for monitoring SpO(2) variations in lambs, and offers new perspectives for researchers interested in continuous monitoring of oxygenation throughout sleep stages and wakefulness.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.