Abstract

A diffusive sampler for formaldehyde originally designed for use in personal monitoring of worker exposure has been evaluated for static measurement of low formaldehyde levels in indoor air. The sampler consists of a 37-mm glass fiber filter impregnated with 2,4-dinitrophenylhydrazine and phosphoric acid and mounted in a polystyrene filter cassette. Formaldehyde is sampled by controlled diffusion and subsequent hydrazone formation on the filter. The hydrazone is eluted from the filter with acetonitrile and analyzed by liquid chromatography and a UV detector. The diffusive sampler was evaluated for static (area) sampling in several industries, offices, and homes with formaldehyde levels of 6-200 ppb. Results from diffusive samplers were compared with results from pumped samplers. Even at low wind velocities (less than 0.01 m.s-1) there was good agreement between pumped and diffusive samples. The sensitivity will be 1 ppb in a 24-h sample, making the sampler especially useful for indoor air monitoring of low formaldehyde levels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.