Abstract

A commercial glucose analyzer, originally designed for monitoring of blood glucose in patients, was tested for use in fermentation processes. The system operates in such a way that the measured value is updated every 90 seconds. The measuring range of the system is 0–5 g glucose/l and the accuracy is ±7%. The response time was found to be approximately 6 min. The system was used to follow fermentations with two different microorganisms, Saccharomyces cerevisiae and Escherichia coli in media containing up to 5 g/l of glucose. The performance was fully satisfactory and the values had a very good correlation with off-line analyses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.