Abstract
Capillary electrophoresis (CE) with contactless conductivity detection (C(4)D) is readily applicable to determinations of amino acids in clinical samples. Most of these analyses employ long separation pathways. This chapter describes CE/C(4)D determinations of 28 biogenic amino acids in a short capillary with an effective length of 18 cm. All the test amino acids can be mutually separated in electrolytes of 0.5-10 mol/L acetic acid. The time of analysis does not exceed 6 min; the limits of detection vary from 0.1 to 1.7 μmol/L for all the analytes. The pretreatment of the biological material is very simple, consisting of the removal of proteins by an addition of acetonitrile and subsequent filtration. The procedure has been successfully applied to determinations of the whole amino acid spectra in blood plasma, urine, saliva, cerebrospinal, and amniotic fluid samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.