Abstract

This report describes the analytical validation and application of the homogeneous mobility shift assay (HMSA) method for the measurement of adalimumab and human antibodies-to-adalimumab (ATA) in serum samples from patients who have lost response to adalimumab treatment. Validation of the ATA- and the adalimumab-HMSA revealed a lower limit of detection to be 0.026U/mL for ATA and 0.018μg/mL for adalimumab in serum samples. Intra-assay and inter-assay precision determination yielded a coefficient of variation of less than 15%, and the accuracy of both assays was within 20%. Adalimumab drug tolerance in the ATA-HMSA was up to 20μg/mL in the test serum. Serum samples from 100 drug-naïve healthy subjects were tested to set-up the cut point of 0.55U/mL for ATA and 0.68μg/mL for adalimumab. Analysis of 100 serum samples from patients who were losing response to adalimumab showed that 26% had an adalimumab level below the cut point, of these 68% were ATA positive. Overall, 44% of the patients (44/100) were positive for ATA. This study presents evidence that drug and anti-drug antibody levels are important determinants of patient response to therapy.