Abstract
This paper reports on the use of a microfluidic dielectric sensor, termed ClotChip, for monitoring the fibrin polymerization effects on whole blood coagulation. The ClotChip sensor readout captures the temporal variation in the normalized real part of the blood dielectric permittivity at 1MHz as it undergoes coagulation. A new readout parameter, S max , is identified in the ClotChip readout curve, which represents the maximum permittivity slope after a permittivity peak is reached. Experiments based on ex vivo treatment of whole blood samples with H-Gly-Pro-Arg-Pro-OH (GPRP) – a fibrin-polymerization inhibitor – demonstrate that S max decreases with increasing GPRP concentrations corresponding to higher levels of fibrin inhibition. Furthermore, S max is shown to exhibit a strong positive correlation (Pearson’s r = 0.883, p < 0.001, n = 12) with the α-angle readout parameter of rotational thromboelastometry – a clinical, viscoelastometry-based, comprehensive assay of blood coagulation. This work shows the ClotChip potential as a point-of-care platform for rapid assessment of fibrin-polymerization defects and their impact on the overall hemostatic status.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.