Abstract

The Bragg peak of protons can be determined by measuring prompt γ-rays. In this study, prompt γ-rays detected by single-photon emission computed tomography with a geometrically optimized collimation system were reconstructed by an iterative method. The falloff position by iterative method (52.48mm) was most similar to the Bragg peak (52mm) of an 80MeV proton compared with those of back-projection (54.11mm) and filtered back-projection (54.91mm) methods. Iterative method also showed better image performance than other methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.